DL8000 Preset Instruction Manual
Revised February-2016 DL8000 Keypad Display E-59
14/03/12 12:26:02
Calibration Failed
Affected
Point Types
The program affects the following point types during calibration:
ZERO EU [104,X,14]
MID point 1 [106,X,15]
MID point 2 [106,X,16]
MID point 3 [106 X,17]
SPAN EU [103,X,18]
Set value [103,X,20]
Calibration mode [103,X,35]
Calibration type [106,X,36]
Units [106,X,37]
Note: The system only checks the point fail alarm during an RTD
calibration.
E.4.4 Calibrating Product Meters
Unlike the DanLoad 6000 (which required manual calibration of
product meters), the DL8000 performs most of the calibration steps
automatically. When appropriate, the program prompts you for required
data. You can enter this information from the keypad display.
The DL8000 provides two parameter-controlled options for selecting
the meter factor [69,X,74] :
Single MF (0)
Multiple MF (Meter factor curve) (1)
Single MF
Regardless of the available number of flow rates, the system supports
only one applicable MF. For example, if the preset has two flow rate set
points (FR1 for low flow rate and FR2 for high flow rate), the system
supports a single MF for both flow rate 1 and 2.