DL8000 Preset Instruction Manual
Revised February-2016 DL8000 Keypad Display E-49
CLEAR
Clr param data (clear the data entry for the
parameter)
BKSP
Clr last char entry
This key clears only the last character and
not the entire value (use CLEAR for that).
ENTER To update value
Right Arrow
Jump to next page (provided the number of
pages is greater than 1)
Left Arrow
Jump to prev page (provided the number of
pages is greater than 1)
Up Arrow Move to Next Line
Down Arrow Move to Prev Line
E.4 Calibration
Warning
When performing calibrations, DO NOT remove a module after you start
the calibration process. The system cannot automatically detect a
module being removed and this may affect the results of your
The keypad display program enables you to calibrate:
Additive Meter
Analog Input
RTD
Product Meter
Caution
During calibration, the DL8000 times-out and reconnects if left idle for
an extended period. In this case, you lose all calibration values and
must reconnect to the DL8000 and restart the calibration. No error or
Each calibration process has its own menu. Select Calibration in
program mode. The Calibration menu displays:
For traversing use
UP | DOWN
LEFT | RIGHT
ENTER to continue
EXIT to Go Back
EXIT to quit
Additive Calib
Analog Calib
RTD Calib
PRODUCT Calib
Select an option to display a calibration menu.
Note: If you select any calibration option, the DL8000 first checks the
status of the W&M switch. If the switch is locked, the DL8000
prevents the calibration and displays a message: