DL8000 Preset Instruction Manual
Revised February-2016 DL8000 Keypad Display E-55
Once the calibration completes, the program displays a summary screen.
You can then accept or reject the new calibration value(s):
14/03/12 12:26:02
Enter ZERO EU
Enter SPAN EU
(ENTER to accept )
Live Dead
-0.09100 0
99.857282 100
(STOP to Reject )
When you press ENTER, the program updates the values and displays a
screen indicating the successful completion of the calibration process.
14/03/12 12:26:02
Calibration Done
The program then redisplays the channel selection screen. You can
begin another AI calibration or exit the AI calibration menu.
If you press STOP, the program restores the old calibration values.
Affected
Point Types
The program affects the following point types during calibration:
ZERO EU [103,X,13]
MID point 1 [103,X,14]
MID point 2 [103,X,15]
MID point 3 [103 X,16]
SPAN EU [103,X,17]
Set value [103,X,19]
Calibration mode [103,X,34]
Calibration type [103,X,35]
E.4.3 Calibrating RTDs
Note: The preset must have an installed RTD module to calibrate
RTDs. If you attempt an RTD calibration without a module, the
system displays the message NO RTD Card in system on the left-
hand screen.
To calibrate RTDs, select the RTD Calib option on the Calibration
menu screen: