DL8000 Preset Instruction Manual
E-56 DL8000 Keypad Display Revised February-2016
14/03/12 12:26:02
For traversing use
UP | DOWN
LEFT | RIGHT
ENTER to continue
EXIT to Go Back
EXIT to quit
Additive Calib
Analog Calib
RTD Calib
PRODUCT Calib
(Pg:1/1)
The preset displays all available RTD channels. In sequence, you select
the appropriate RTD channel, number of mid-points, and the units for
calibration:
14/03/12 12:26:02
Sel RTD Ch
Sel No of Mid pt
Sel Unit
Zero UE Value
Span EU Value
(SELECT)
RTD Ch 1
Once you select an RTD channel, the program shows the existing zero
and span values of the selected channel. You then enter the number of
mid-points and the units:
14/03/12 12:26:02
Sel RTD Ch
Sel No of Mid pt
Sel Unit
Zero UE Value
Span EU Value
(SELECT)
RTD Ch 1
Zero
100.000000
229.000000
14/03/12 12:26:02
Sel RTD Ch
Sel No of Mid pt
Sel Unit
Zero UE Value
Span EU Value
(SELECT)
RTD Ch 1
Zero
Ohms
100.000000
229.000000
Point Fail Alarm
Note: If a point fail alarm occurs for the selected channel, the system
displays a Point Fail alarm message at the bottom of the right-
hand screen (as shown above). This alarm does not stop the
calibration; it is provided for information only. During
calibration the program clears the message if the condition does
not exist.