DL8000 Preset Instruction Manual
E-56 DL8000 Keypad Display Revised February-2016
For traversing use
UP | DOWN
LEFT | RIGHT
ENTER to continue
EXIT to Go Back
EXIT to quit
Additive Calib
Analog Calib
RTD Calib
PRODUCT Calib
The preset displays all available RTD channels. In sequence, you select
the appropriate RTD channel, number of mid-points, and the units for
calibration:
Sel RTD Ch
Sel No of Mid pt
Sel Unit
Zero UE Value
Span EU Value
RTD Ch 1
Once you select an RTD channel, the program shows the existing zero
and span values of the selected channel. You then enter the number of
mid-points and the units:
Sel RTD Ch
Sel No of Mid pt
Sel Unit
Zero UE Value
Span EU Value
RTD Ch 1
Zero
100.000000
229.000000
Sel RTD Ch
Sel No of Mid pt
Sel Unit
Zero UE Value
Span EU Value
RTD Ch 1
Zero
Ohms
100.000000
229.000000
Note: If a point fail alarm occurs for the selected channel, the system
displays a Point Fail alarm message at the bottom of the right-
hand screen (as shown above). This alarm does not stop the
calibration; it is provided for information only. During
calibration the program clears the message if the condition does
not exist.