154
Reference Manual
00809-0100-4026, Rev KC
Operation
October 2018
Operation
LCD display parameters
Table 6-1. LCD Display Parameters and Presentation on Display
6.2.3 Viewing measurement data in RRM
To view measurements, such as level, signal strength, etc. in RRM, select the Tools > Device Display
option and select the Level tab:
Figure 6-4. Presentation of Measurement Data in RRM
Parameter
Presentation on
display
Description
Level LVL Product level
Distance DST Distance from the upper reference point to the product
surface
Level Rate LR The speed of level movement up or down
Signal Strength AMP The signal amplitude of the surface echo
Volume Only measurement
unit is shown
Total product volume
Internal Temperature ITEMP Temperature inside the transmitter housing
AOut Current ANOUT Analog Output 4 -20 mA current
Percent of Range % RNG Level value in percent of total measurement range
Comm Quality COM Q Digital communication signal quality
Signal Quality SIG Q The signal quality
Surface/Noise
Margin
SNM The relationship between the surface peak amplitude and the
strongest noise peak amplitude