Loading...
Product Certifications
249
Product Certifications
October 2018
Reference Manual
00809-0100-4026, Rev KC
N3 China Type N
Certificate: NEPSI GYJ17.1421X
Standards: GB3836.13-2013, GB3836-15-2000,
GB3836.16:2008, GB3836.18:2010,
GB50257-2014
Markings: Ex nA IIC T4 Gc
Ex ic IIC T4 Gc
IP66/IP67
Specific conditions for safe use (X):
1. See certificate for specific conditions.
B.11 Technical Regulations
Customs Union (EAC)
EM Technical Regulations Customs Union (EAC) Flameproof
Certificate: RU C-SE.AA87.B.00802
Markings: Ga/Gb Ex db ia IIC T4 X, (-40°C Ta
+60°C/+70°C)
Ex ta IIIC T69°C Da X, (-40°C Ta +70°C)
Ex ta IIIC T79°C Da X, (-40°C Ta +70°C)
Specific Conditions for Safe Use (X):
1. See certificate for specific conditions.
IM Technical Regulations Customs Union (EAC) Intrinsic
Safety
Certificate: RU C-SE.AA87.B.00802
Markings: 0Ex ia IIC T4 Ga X, (-50°C Ta +60°C/+70°C)
Ga/Gb Ex ib IIC T4 X, (-50°C Ta
+60°C/+70°C)
Ex ia IIIC T79°C Da X, (-50°C Ta +70°C)
Ex ia IIIC T69°C Da X, (-50°C Ta +60°C)
Da/Db Ex ib IIIC T79°C Da X, (-50°C Ta
+70°C)
Da/Db Ex ib IIIC T69°C Da X (-50°C Ta
+60°C)
Specific Conditions for Safe Use (X):
1. See certificate for specific conditions.
B.12 Japan
E4 Flameproof
Certificate: CML 17JPN1334X
Markings: Ex d ia IIC T4 Ga/Gb
(-40 °C Ta +60 °C/+70 °C)
Specific conditions for safe use (X):
1. See certificate for specific conditions.
B.13 India
Flameproof
Certificate: P392482/1
Markings: Ex db ia T4 Ga/Gb
Ex ia T4 Ga
Specific conditions for safe use (X):
1. See certificate for specific conditions.
B.14 Ukraine
Flameproof, Intrinsically Safe
Certificate: UA.TR.047.C.0352-13
Markings: 1 Ex de IIC T4X
1 Ex de ib ia IIC T4 X
1 Ex de ia IIC T6 X
Specific Conditions for Safe Use (X):
1. See certificate for specific conditions.
B.15 Uzbekistan
Safety (import)
Certificate: UZ.SMT.01.342.2017121
U
i
I
i
P
i
C
i
L
i
Maximum input
parameters for
Ex ic HART
42.4 V DC 23 mA 1 W 7.25 nF 0
Maximum input
parameters for
Ex ic Fieldbus
32 V DC 21 mA 0.7 W 4.95 nF 0
U
i
I
i
P
i
C
i
L
i
Entity parameters
HART
30 V 130 mA 1 W 7.26 nF 0 mH
Entity parameters
Fieldbus
30 V 300 mA 1.5 W 4.95 nF 0 mH