Service and Troubleshooting
October 2018
Service and Troubleshooting
7.14.6 Volume calculation status
Volume calculation status messages that may appear on the integral display, on the Field Communicator,
or in the RRM program are shown in Ta ble 7 -6.
Table 7-6. Volume Status
Message Description Action
Level is below lowest
strapping point
The measured level is below the
lowest point in the given strapping
table.
For a correct volume calculation in this
region, change the strapping table.
Level is above highest
strapping point
The measured level is above the
highest point in the given strapping
table.
For a correct volume calculation in this
region, change the strapping table.
Level out of range The measured level is outside the
given tank shape.
Check if the correct tank type is chosen,
and check the configured Tank Height.
Strap table length not
valid
The configured strap table length is
too small or too large.
Change the strapping table size to a valid
number of strapping points. A maximum
number of 20 strapping points can be
entered.
Strap table not valid The strapping table is not correctly
configured.
Check that both level and volume values
in the strapping table are increasing with
strapping table index.
Level not valid The measured level is not valid. No
volume value can be calculated.
Check “Measurement Status” on
page 183, “Warning Messages” on
page 183, and “Error Messages” on
page 182.
Volume configuration
missing
No volume calculation method is
chosen.
Configure Volume.
Volume not valid The calculated volume is not valid. Check the other volume status messages
for the reason.