Emerson Process Management GmbH & Co. OHG 4-13
4
Maintenance
X-STREAM X2FD
Instruction Manual
HASXMDE-IM-EX
01/2015
Fig. 4-8: Converter lling sequence
4.3 Vapor Recovery Application - Special Information
5. Remove the end cap.
6. Behind the washer, the vessel is lled with
converter material. Remove the washer.
7. Remove this old material and ll the vessel
with new converter material, considering
to
hold the vessel vertically while lling
• completely ll the vessel
keep the inner pipe free of converter
material.
8. Put the washer on top of the vessel / new
material.
9. Install the end cap.
10. Insert the converter vessel into the jacket
and place it into the assembly.