EasyManuals Logo

Emerson X-STREAM XE User Manual

Emerson X-STREAM XE
536 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #196 background imageLoading...
Page #196 background image
Emerson Process Management GmbH & Co. OHG6-34
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
Ch1
Gases
Zero Gas: 0.000 ppm
Span Gas: 50000.000 ppm
Range Gases..
Current Range: Range 1
6.2.3 Setup Menu
6.2.3.2.1 Setup Calibration Gases
6.2.3.2.1.1 Setup Range Gases
Enter this line to setup the zero gas concen-
tration.
Enter this line to setup the span gas concen-
tration.
Submenu to specify different calibration gases
for several ranges (if instrument is setup to
use different ranges);
below.
Information about currently used range.
This menu enables to specify separate zero
and span calibration gas concentrations for
each range of the selected channel.
Note!
The calibration gases units are as setup for
the currently selected channel;
6-26.
Note!
The calibration gases units are as setup for
the currently selected channel; 6-26.
Setup..
Calibration..
Gases..
Setup..
Calibration..
Gases..
Range Gases..
Ch1
Range Gases
Zero Gas1: 0.000 ppm
Zero Gas2: 0.000 ppm
Zero Gas3: 0.000 ppm
Zero Gas4: 0.000 ppm
Span Gas1: 500.000 ppm
Span Gas2: 5000.000 ppm
Span Gas3: 25000.000 ppm
Span Gas4: 50000.000 ppm

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson X-STREAM XE and is the answer not in the manual?

Emerson X-STREAM XE Specifications

General IconGeneral
BrandEmerson
ModelX-STREAM XE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals