EasyManuals Logo

Emerson X-STREAM XE User Manual

Emerson X-STREAM XE
536 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #207 background imageLoading...
Page #207 background image
Emerson Process Management GmbH & Co. OHG 6-45
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
6
Software Menus
Ch1
Ranges
Current Range: Range1
Range Start: 0.000 ppm
Range End: 50000.000 ppm
Starts & Ends..
Range Control: Manual
Auto Ranging..
6.2.3.3.1 Setup Measurement Ranges
6.2.3 Setup Menu
Specify the limits for up to four ranges per
channel in one single menu, making it easier
to adjust several limits at a glance;
6-46.
Select the range to be congured.
Available options: Range1 ... Range4
Enter this menu to congure up to four ranges
per channel.
For the current range, specify start and end
concentration limits.
For the current range, specify how range
switching is done.
Available options:
Manual, Remote, Automatic
Note!
Selecting Remote or Automatic range con-
trol is not possible, if identical ranges end
values are specied (
6-46)!
Specify the switchover levels for up to four
ranges per channel;
6-46.
Setup..
Measurement..
Ranges..

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson X-STREAM XE and is the answer not in the manual?

Emerson X-STREAM XE Specifications

General IconGeneral
BrandEmerson
ModelX-STREAM XE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals