EasyManua.ls Logo

Emerson X-STREAM XE - Page 206

Emerson X-STREAM XE
536 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Emerson Process Management GmbH & Co. OHG6-44
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
Ch1
Measurement 1of2
Ranges..
Damping..
Linearization..
Cross Interference..
Average..
Delay..
Ch1
Measurement 2of2
Cut-off: None
Pressure Compensation ..
6.2.3.3 Setup Measurement
6.2.3 Setup Menu
Multi-channel unit:
In SELECT COMPONENT, select the chan-
nel to be congured.
Multi-channel unit:
Press
left to enter SELECT COMPONENT, to
change the settings for another channel.
If the system is setup accordingly, the access
code for level 2 must be entered to gain ac-
cess to this menu.
Submenu to congure up to four ranges per
channel;
6-45.
Setup t
90
times;
6-48.
Setup and activate linearization;
6-55.
Setup and activate cross interference com-
pensation;
6-55.
Setup and activate averaging;
6-58.
Setup delay time;
6-59.
Cut-off mode: Output values are limited, in
case they exceed the congured range limits,
or are becoming negative.
Available options:
None: cut-off mode is disabled
RngLimits: values are limited to range limits
NonNegat.: negative values are output as '0'
Setup..
Measurement..
Note!
Cut-off always is disabled during calibra-
tions!
Setup ambient pressure for compensation;
6-60