6 • SET-UP
Art.-Nr.: 80116-271
04.06
APC
AutoPurge
V 1.4.x The instrument is purged with gas automatically when it is
plugged into the APC receptacle and an instrument that is
already plugged into the APC receptacle is purged with gas auto
-
matically when the unit is started up.
APC Purge
Duration
V 1.4.x Selection of time for which the instrument is purged with gas
automatically:
0 to 10 s.
APC
PurgeFlow
V 1.3.x Only in V 1.3.x: The instrument is purged with gas automatically
when it is plugged into the APC receptacle and an instrument that
is already plugged into the APC receptacle is purged with gas
automatically when the unit is started up.
As from V 1.4.x: Selection of purge flow (in %) at which the instru-
ment is purged with gas automatically. Purge flow relates to the
default COAG-Flow setting stored in the instrument.
max. APC
cyl.pres
-
sure
V 1.5.x Setting of maximum cylinder pressure in the argon cylinder used.
Correct reporting on the cylinder level display on the HF surgical
device depends upon the maximum cylinder pressure setting of
the actual argon gas bottle used:
100 to 240 bar.
Sound
sample
V 1.2.x Selection of type of warning signals.
DRY ° /
SWIFT
°
V 1.3.x to
V
1.4.x
Only relevant for VIO 300 D:
ON:
The DRY CUT °/SWIFT COAG ° modes are used.
OFF:
The DRY CUT/SWIFT COAG modes are used.
SWIFT ° V 1.5.x Only relevant for VIO 300 D:
ON:
The SWIFT COAG ° mode is used.
OFF:
The SWIFT COAG mode is used.
DRY ° V 1.5.x Only relevant for VIO 300 D:
ON:
The DRY CUT ° mode is used.
OFF:
The DRY CUT mode is used.
Setting Avaiable from Description