9 • Troubleshooting
66 / 144
Art.-Nr.: 80116-271
04.06
B 13 1 => Error file 1 Errors in list management:
Occurs if an error is identi
-
fied in the corresponding
flash file when managing
the event or error list.
Delete lists. If this recurs,
there could be a fault in the
flash memory => Replace
assembly. Inform ERBE
Tübingen Technical Ser
-
vice.
2 => Error file 2
10 => Event file 1
20 => Event file 2
B 14 CRC error:
Occurs if an error is identi-
fied by the continuous CRC
monitoring of the program.
(e.g. A bit flipping in the
flash memory or an error
when downloading soft
-
ware).
Reload system software.
If this recurs, there could be
a fault in the flash memory
=> Replace assembly.
Inform ERBE Tübingen
Technical Service
B 15 FLASH RAM verification
error:
Occurs if an error is identi-
fied by the continuous mon-
itoring of the program. (e.g.
a bit flips in DRAM mem
-
ory).
One-off event -> EMC prob-
lem => Check environment.
Reproducible -> Error in
system memory =>
Replace assembly. Inform
ERBE Tübingen Technical
Service.
B 16 Insufficient EEPROM mem-
ory:
Occurs if the memory for
user programs is full.
Delete programmes that
are not required.
B 17 CAN ID of second
activation signal.
Double activation:
Occurs if two activation sig-
nals (e.g. both pedals of a
footswitch) are present
simultaneously (within 100
ms).
User error or faulty activa-
tion element.
B 18 Bit combination of HF
module and APC
module:
1000 or 100 HF mod-
ule (Coag)
2000 or 200 HF mod-
ule (Cut)
4000 or 400 APC
module
No deactivation signal:
Occurs if a module involved
in activation (HF module or
APC module) fails to react
to a deactivation request for
longer than 110 ms.
Communication error.
One-off event -> EMC prob-
lem => Check environment.
Reproducible -> Software
error => Inform ERBE
Tübingen Technical Ser
-
vice.
Recognizing module: B = Control panel
Recog-
nizing
module
Error
code
Additional informa-
tion
Description Remedy