67 / 144
9 • Troubleshooting
Art.-Nr.: 80116-271
04.06
B 19 PowerFail:
Occurs if a PowerFail signal
is received but the Power
-
Fail does not actually hap-
pen within 2 s.
Check house installation.
Check low-voltage power
supply unit.
B 1A Number of the incor-
rect parameter (Hex):
0: Activation assign-
ment
3: CutMode
4: CutIntensity
5: CutEffect
6: CoagIntensity
7: CoagEffect
8: CoagMode
B: CutAPCFlow
C: CoagAPCFlow
D: CutIESFlow
E: CoagIESFlow
F: IES-Basic Suction
10: IES-Run-On Time
Error in parameter memory:
Occurs if an error is identi
-
fied by the checksum
parameter monitoring of a
program. (e.g. a bit flips in
DRAM memory).
One-off event -> EMC prob-
lem => Check environment.
Frequent event -> Error in
system memory =>
Replace assembly.
Reproducible -> Software
error => Inform ERBE
Tübingen Technical Ser
-
vice.
B 1B CRC check not yet com-
pleted: Occurs if the user
wants to operate the unit
after downloading software
before there has been at
least one successful CRC
check.
Wait until the CRC check
has been completed.
B 1C The maximum ON time has
been exceeded: Occurs if
activation lasts longer than
the ON time selected in
SET-UP.
User error.
B 1D Number of the incor-
rect parameter (Hex):
101 Cut-Mode
102 Coag-Mode
201 Cut-Effect
202 Coag-Effect
301 Cut-Power
302 Coag-Power
501 Cut-APC-Flow
502 Coag-APC-Flow
Incorrect instrument param-
eters:
Occurs if an instrument is
detected via instrument
recognition and the param
-
eter is outside the permissi-
ble tolerances.
Check instrument software
version. Replace instru
-
ment. Check IIF.
Recognizing module: B = Control panel
Recog-
nizing
module
Error
code
Additional informa-
tion
Description Remedy