9 • Troubleshooting
68 / 144
Art.-Nr.: 80116-271
04.06
B 1E Button code:
0x001
=> buttonUp
0x002 => buttonDwn
0x004 => buttonEnter
0x008 =>
buttonReceptacle
1
0x010
=>
buttonReceptacle
2
0x020 =>
buttonReceptacle
3
0x040 => buttonNE
0x080 =>buttonAPC1
0x100 =>buttonAPC2
Keyboard error:
Occurs if a button pressed
is recognized during initial
-
ization.
User error or
=> Check keyboard.
B 1F NESSY symmetry
value in %.
NESSY symmetry monitor-
ing:
Occurs if an error is sig-
naled by the Nessy symme-
try monitoring during
activation.
User error.
Should be certain that the
measured value is incorrect
=> Check NE module.
B 21 Incorrect or missing bit map
file:
Occurs if an incompatible
bit map file is found.
Transfer compatible bit map
file.
B 22 Occurs if activation was ter-
minated automatically (e.g.
by AutoStop) and tissue is
still touched.
Discontinue touching.
B 23 Configuration error of HF
CPU: Occurs if the HF CPU
transmits an invalid status
message (0X27).
Check software version of
HF CPU.
B 24 Occurs if an incorrect sta-
tus message is received
from the IES
2.
Check software status of
the system components.
If the error is reproducible
notify => Technical Service
ERBE Tübingen.
Recognizing module: B = Control panel
Recog-
nizing
module
Error
code
Additional informa-
tion
Description Remedy