9 • Troubleshooting
Art.-Nr.: 80116-271
04.06
B 91 No status message from
IES
module:
Occurs if the IES module is
disconnected from the sys
-
tem.
Check connectors. Check
IES
2.
B 92 IES module recognized:
Occurs if an IES
module is
detected.
Information.
B 93 Multifunctional footswitch
recognized:
Occurs if a multifunctional
footswitch is connected.
Information.
B 94 Multifunctional footswitch
has been disconnected
from the system:
Occurs if a multifunctional
footswitch is disconnected
from the system.
Information.
B 95 Number of output
receptacle
*0X10000
+ instru-
ment number.
New instrument recog-
nized by system:
Occurs if an instrument with
instrument recognition is
connected.
Information.
B 97 Error in the program modifi-
cation list:
Occurs if the checksum of a
stored setting is not correct
(e.g. due to a storage cell
flipping). The setting is lost
and the basic setting for
this test program is
restored.
One-off event -> Interfer-
ence => Check ambient
condition.
Frequent event -> Memory
error => Replace assembly.
Reproducible -> Software
error => Inform ERBE
Tübingen Technical Ser
-
vice.
B 98 Error in the program list:
Occurs if the checksum of a
stored program is not cor
-
rect (e.g. due to a storage
cell flipping). The test pro
-
gram is lost.
One-off event -> Interfer-
ence => Check ambient
condition.
Frequent event -> Memory
error => Replace assembly.
Reproducible -> Software
error => Inform ERBE
Tübingen Technical Ser
-
vice.
Recognizing module: B = Control panel
Recog-
nizing
module
Error
code
Additional informa-
tion
Description Remedy

Table of Contents