9 • Troubleshooting
72 / 144
Art.-Nr.: 80116-271
04.06
B 99 Activation type: This activation type is not
available:
Occurs if the user assigns
an activation signal that is
not currently available. (e.g.
unconnected footswitch).
User error or activation ele-
ment (see additional infor-
mation) faulty and therefore
not available on the system.
1 => Dual-pedal foot-
switch both pedals
2 => Dual-pedal foot-
switch only Coag
3 => Dual-pedal foot-
switch only Cut
4 => Single-pedal
footswitch
5 => AutoStart 1
6 => AutoStart 2
B 9A Please check the clock:
Occurs if the VIO was
switched off for such a long
time that the supply current
to the built-in realtime clock
was no longer sufficient.
Set time.
If frequent event, check
BT10.
B 9B Master remote control rec-
ognized:
Occurs if a master remote
control is recognized by the
VIO.
Information.
B 9C Master remote control has
been disconnected from the
system:
Occurs if the master remote
control is disconnected
from the system.
Information.
B 9D Remote control recognized:
Occurs if a remote control
is recognized by the VIO.
Information.
B 9E Remote control has been
disconnected from the sys
-
tem:
Occurs if the remote control
is disconnected from the
system.
Information.
Recognizing module: B = Control panel
Recog-
nizing
module
Error
code
Additional informa-
tion
Description Remedy