9 • Troubleshooting
Art.-Nr.: 80116-271
04.06
B 9F Number of output
receptacle.
Instrument has been dis-
connected from the system:
Occurs if an instrument with
instrument recognition is
disconnected from the sys
-
tem.
Information.
B A0 Number of output
receptacle.
No other mode can be set
for this instrument:
Occurs if the user wants to
assign a different mode to
an instrument with a fixed
mode.
Information.
B A1 No status message from
HF module:
Occurs if the control panel
does not receive a status
message from the HF mod
-
ule (for longer than 1 s).
One-off event
-> Interference
=> Check
ambient conditions.
Frequent event
-> Memory error
=> Check CPU+sensors.
B A3 Error number of foot-
switch (see error
description for foot
-
switch):
+
0x100 (with single-
pedal footswitch)
+
0x200 (with dual-
pedal footswitch)
+
0x400 (with multi-
functional footswitch)
Footswitch not assigned:
Occurs if a footswitch which
has not been assigned to
an output socket is
pressed.
User error.
B A4 CAN ID of second
footswitch or 2 if it
cannot be assigned to
any CAN ID.
Two footswitches are con-
nected: Occurs if two foot-
switches of the same type
are connected.
User error.
B A6 EEPROM is updated:
Occurs if the unit is pro
-
grammed using a con-
nected PC.
Information.
B A8 CAN ID of module. Invalid ECB version:
Occurs if a module with an
invalid ECB version logs
on.
Check software version of
system components.
Recognizing module: B = Control panel
Recog-
nizing
module
Error
code
Additional informa-
tion
Description Remedy