9 • Troubleshooting
74 / 144
Art.-Nr.: 80116-271
04.06
B A9 Activation not possible until
settings are confirmed:
Occurs if the user wants to
perform activation after
switching on the unit with
-
out confirming the settings
beforehand.
User error.
B AA CAN ID of activation
request.
No valid mode is assigned:
Occurs if an output channel
is activated without a mode
being assigned to it.
User error. If necessary,
load upgrade.
B AB CAN ID of activation
request.
Activation is only possible
with a valid instrument:
Occurs if an MF receptacle
is activated at which no
instrument is recognized.
User error or check plug-in
contacts. Check instrument
software.
B AC 0x140 (CAN ID of
AutoStart monitor).
AutoStart is not possible if
contact is made during
assignment:
Occurs if the AutoStart
function is assigned while
contact has already been
recognized.
User error or check AUTO
START monitor (CPU+sen
-
sors).
B AD CAN ID of module. Error with version check:
Occurs if a module with an
invalid version tries to log
on.
Check software version of
system components.
B AE No modules logged on:
Occurs if no modules log on
during the initialization
phase.
Check connectors, ECB
and electrical supply.
B AF NE module ceases to
respond:
Occurs if the NE module
ceases responding.
Check NE module.
B BO NESSY symmetry
value in
%.
NESSY symmetry warning:
Occurs if the Nessy sym-
metry monitoring signals a
value between 20
% and
50
%.
User error.
It should be certain that the
measured value is incorrect
=> Check NE module.
Recognizing module: B = Control panel
Recog-
nizing
module
Error
code
Additional informa-
tion
Description Remedy