75 / 144
9 • Troubleshooting
Art.-Nr.: 80116-271
04.06
B B1 NESSY current value. NESSY current density
warning:
Occurs if the Nessy current
density monitoring signals
a value above the limit
curve.
User error.
It should be certain that the
measured value is incorrect
=> Check NE module.
B B2 IES footswitch detected:
Occurs if an IES
footswitch
is connected.
Information.
B B3 Calibration of keyboard:
Occurs if the capacitive
keyboard is recalibrated.
Information.
If frequent event, replace
assembly.
B B4 Rinsing of an APC instru-
ment:
Occurs if rinsing of an APC
instrument is initiated.
Information.
B B5 CAN warning:
Occurs if the error counter
of the CAN controller
reaches a critical value.
ECB communication error.
Check ambient conditions.
B B6 Incorrect receptacle version
signal:
Occurs if an incorrect
receptacle version signal is
received.
Check software version of
receptacle module.
B B7 Permissible limit value of
AUTO START activation
has been exceeded:
Occurs if an attempt is
made to assign AUTO
START activation a non-
permissible power limitation
value.
Information.
B B8 New interface module rec-
ognized:
Occurs if an interface mod-
ule (e.g. VIO PORTAL) logs
on to the system.
Information.
Recognizing module: B = Control panel
Recog-
nizing
module
Error
code
Additional informa-
tion
Description Remedy