9 • Troubleshooting
76 / 144
Art.-Nr.: 80116-271
04.06
B B9 Interface module has been
disconnected from the sys
-
tem:
Occurs if an interface mod-
ule (e.g. VIO PORTAL) is
disconnected from the sys
-
tem.
Information.
B BA Warning from interface
module:
Occurs if an interface mod-
ule (e.g. VIO PORTAL) sig-
nals an error.
Check interface module.
B BB Reminder that a safety
check is due:
Occurs when a safety
check is due.
Information.
B BC VEM 2 module recognized:
Occurs if receptacle exten
-
sion module VEM 2 logs
into the system.
Information.
B BD VEM 2 module discon-
nected:
Occurs if receptacle exten-
sion module VEM 2 is dis-
connected from the system.
Information.
B BE VEM 2 module receptacle 1
no longer ready for opera
-
tion:
Occur if the 1st receptacle
of receptacle extension
module VEM
2 no longer
responds.
Check receptacle 1 in
VEM
2 module.
B BF VEM 2 module receptacle 2
no longer ready for opera
-
tion:
Occur as if the 2nd recepta-
cle of receptacle extension
module VEM
2 no longer
responds.
Check receptacle 2 in
VEM
2 module.
Recognizing module: B = Control panel
Recog-
nizing
module
Error
code
Additional informa-
tion
Description Remedy