77 / 144
9 • Troubleshooting
Art.-Nr.: 80116-271
04.06
B CO Neither footswitch nor
AUTO START assigned:
Occurs if an instrument
without a fingerswitch is
recognized and neither a
footswitch nor AUTO
START is assigned.
Information.
B C1 EIP 2 was recognized by
the system:
Occurs if the EIP 2 is con-
nected to the system.
Information.
B C2 EIP 2 was disconnected
from the system:
Occurs if EIP 2 is discon-
nected from the system.
Information.
B C3 IES 2 footswitch was dis-
connected from the system:
Occurs if the IES
2 foot-
switch is disconnected from
the system.
Information.
B C4 Purge function not
assigned:
Occurs if the purge button
on the APC is pressed and
no APC receptacle was
selected.
Select APC receptacle.
B FC Power Down:
Occurs if the unit is
switched off.
Information.
B FD System Reset:
Occurs if a system reset is
performed (e.g. when
switching on unit).
Information.
B FE PowerFail:
Occurs if a PowerFail is sig-
naled (e.g. when switching
off unit).
Information.
Recognizing module: B = Control panel
Recog-
nizing
module
Error
code
Additional informa-
tion
Description Remedy