9 • Troubleshooting
80 / 144
Art.-Nr.: 80116-271
04.06
C 26 Measurement values of
P_HF and P_NT do not
agree.
E.g. HF power output > NT
power output.
Check power supply unit
(TP power supply unit).
Check generator
(TP
generator).
Check calibration.
C 30 Redundant measurement
value for HF voltage too
great during activation.
Check generator
(TP
generator).
Check sensors.
C 31 Measurement value for HF
voltage too great with OFF.
Check power supply unit
(TP power supply unit).
Check generator
(TP
generator).
Check sensors.
C 32 Measurement value for HF
voltage too small with OFF.
Check power supply unit
(TP power supply unit).
Check generator
(TP
generator).
Check sensors.
C 33 Measurement value for HF
voltage too great with ON.
Check power supply unit
(TP power supply unit).
Check generator
(TP
generator).
Check sensors.
C 34 Measurement value for HF
voltage too small with ON.
Check power supply unit
(TP power supply unit).
Check generator
(TP
generator).
Check sensors.
C 35 Measurement value for HF
current too great with OFF.
Check power supply unit
(TP power supply unit).
Check generator
(TP
generator).
Check sensors.
Recognizing module: C = CPU + Sensors
Recog-
nizing
module
Error
code
Additional infor-
mation
Description Remedy