81 / 144
9 • Troubleshooting
Art.-Nr.: 80116-271
04.06
C 36 Measurement value for HF
current too small with OFF.
Check power supply unit
(TP power supply unit).
Check generator
(TP
generator).
Check sensors.
C 37 Measurement value for HF
current too great with ON.
Check power supply unit
(TP power supply unit).
Check generator
(TP
generator).
Check sensors.
C 38 Measurement value for HF
current too small with ON.
Check power supply unit
(TP power supply unit).
Check generator
(TP
generator).
Check sensors.
C 41 Measurement value for
spark too great with OFF.
Check generator
(TP
generator).
Check sensors.
C 42 Measurement value for
spark too small with OFF.
Check generator
(TP
generator).
Check sensors.
C 43 Measurement value for
spark too great with ON.
Check generator
(TP
generator).
Check sensors.
C 44 Measurement value for
spark too small with ON.
Check generator
(TP
generator).
Check sensors.
C 45 Measurement value for
phase too great with OFF.
Check generator
(TP
generator).
Check sensors.
C 46 Measurement value for
phase too small with OFF.
Check generator
(TP
generator).
Check sensors.
Recognizing module: C = CPU + Sensors
Recog-
nizing
module
Error
code
Additional infor-
mation
Description Remedy