9 • Troubleshooting
88 / 144
Art.-Nr.: 80116-271
04.06
2,3,5,6
(socket)
82 Short circuit activation rec-
ognition R_NEU: The resis-
tor value determined for the
activation recognition cor
-
responds to a short circuit.
Check instrument and
socket module.
2,3,5,6
(socket)
83 No-load activation recogni-
tion R_NEU: The resistor
value determined for the
activation recognition cor
-
responds to no-load.
Check instrument and
socket module.
2,3,5,6
(socket)
84 Instrument recognition R-
window violated: The resis
-
tor value determined for the
instrument recognition can
-
not be reliably assigned to
a setpoint.
Check instrument and
socket module.
2,3,5,6
(socket)
85 Short circuit activation rec-
ognition R_ALT: The resis-
tor value determined for the
activation recognition cor
-
responds to a short circuit.
Check instrument and
socket module.
2,3,5,6
(socket)
86 Activation error: Activation
not consistent with instru
-
ment type.
Check instrument and
socket module.
2,3,5,6
(socket)
87 Framing error: The UART
has not detected a positive
bit (stop bit) at the end of
the received byte.
Check serial interface
IC33, IC34 and wiring.
2,3,5,6
(socket)
88 Parity error: The check for
even parity was unsuc
-
cessful.
Check serial interface
IC33, IC34 and wiring.
2,3,5,6
(socket)
89 Unknown identifier: The
Interrupt routine has
detected a non-agreed
identifier.
IIF FW version compatible
with CPU + sensors?
Check serial interface
IC33, IC34 and wiring.
2,3,5,6
(socket)
8A DS2430 CRC error: A data
record received from
instrumentation is recog
-
nized as being defective.
Check instrument and
socket module.
Recognizing module: 2,3,5,6 = IIF (Instrument Interface) of corresponding receptacle slot
Recog-
nizing
module
Error
code
Additional infor-
mation
Description Remedy