EasyManua.ls Logo

Erbe VIO 200 D - Page 89

Erbe VIO 200 D
144 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
89 / 144
9 • Troubleshooting
Art.-Nr.: 80116-271
04.06
2,3,5,6
(socket)
8B Non-permissible instru-
ment recognition R_S3:
Non-permissible value
determined for resistor-
coded instrument recogni
-
tion.
Carry out reset. Repro-
gram µC. Replace µC.
2,3,5,6
(socket)
8C Non-permissible sockets
and relay status:
Non-permissible value
determined for socket rec
-
ognition.
Carry out reset. Repro-
gram µC. Replace µC.
2,3,5,6
(socket)
8D Non-permissible instru-
ment recognition R_S4:
Non-permissible value
determined for resistor-
coded instrument recogni
-
tion.
Carry out reset. Repro-
gram µC. Replace µC.
2,3,5,6
(socket)
8E Non-permissible instru-
ment recognition R_S8:
Non-permissible value
determined for resistor-
coded instrument recogni
-
tion.
Carry out reset. Repro-
gram µC. Replace µC.
2,3,5,6
(socket)
8F UART timeout:
UPR "Receive_UART" has
not received anything
351us after call-up.
Check serial interface
IC33, IC34 and wiring.
2,3,5,6
(socket)
90 DS2430_Write error:
Attempt to write to
EEPROM for instrument
recognition unsuccessful.
Check instrument and
socket module.
Recognizing module: 2,3,5,6 = IIF (Instrument Interface) of corresponding receptacle slot
Recog-
nizing
module
Error
code
Additional infor-
mation
Description Remedy

Table of Contents

Other manuals for Erbe VIO 200 D

Related product manuals