F20
Keyboard Lock
** Off (Disable)
On (Enable)
When keyboard lock is = On, only [Zero], [Tare], [Set] & [On/Off] keys are accessible during operation status.
F21
Weigh & ATM Mode
Lab 2 Output Print
Format
** STD
CUSTOM
F22
Piece Count Mode
Lab 2 Output Print
Format
** STD
CUSTOM
F24
Animal Weighing
Mode Lab 2 Output
Print Format
** STD
CUSTOM
F25
Check Mode
Mode 0
** Mode 1
Mode 0 = Static Check Mode (only stable weight results are checked). If selected, input also the delay time interval.
Delay time interval is the time value (in second) from when a stable result is detected, and before the corresponding check result
signal/output is triggered.
Mode 1 = Standard/Dynamic Check Mode (real-time check result is given disregarding weight results are stable or not). If Mode 1 is used
for dynamic application, set also F26.
F26
Near Zero Value
** 000000
Near Zero value is used for dynamic weight check applications to bypass unnecessary Lo signals during uploading and unloading process.
Notes: -
Value inputted valid only when Check function is activated.
No check results will be given to weight results which is less than the near zero value.
Near zero weight value can be any value between 20d and Lo limi
Any near zero value which less than 20d will be ignored. Instrument deems 20d as minimum near zero weight value.
F28
Ask for Operator
Number when Power
On
Yes
** No