Yes: - Instrument will ask for operator number input when power on. Input a 4-digit operator number or press [CE/x10] to skip the operator
number when being asked. If a different operator number is inputted, all accumulated memory will be erased.
No: - Instrument will not ask for operation number when instrument power on.
The operator number inputted will be outputted to printout formats Lab 3 ~ Lab 5.
Note: - If F11 Note is set to On, F28 will be forced to Yes automatically.
F29
Read Calibration & Parameter Counts and Pointer Value
C (Calibration count): - displays total times of calibration.
O (Parameter set count): - displays total times that the essential parameters (F80~F88) have been altered.
Press [Zero] to quit to F29 or [Check] to display the current pointer value. Refer to the operation manual for more details about Pointer Value.
F30
Letters and Symbols
for Customer &
Product Code
Manual Entry via
Keyboard on
Instrument
** On
Off
Note: - When a USB external keyboard is plugged in, setting will be forced to On when powered on.
F32
External Input
Assignment
in1
(Input 1)
in2
(Input 2)
in3
(Input 3)
1 of the below 9 key function can be assigned to each individual external input.
1. ZEro = [Zero]
2. tAre = [Tare]
3. Print = [Print/M+]
4. ACC-Mr = [MR]
5. CLEAr = [CE/x10]
6. unit = [Unit]
7. Set = [Set]
8. Check = [Check]
9. FunC = [Func]
To assign a specific key function to external inputs: -
1. Press [Print/M+].