FT-50 Fill Level Inspector Service Manual
6-28
19. To store the settings in system memory, close dip switch S1-2
and then press Key 5 until SAVE DEFAULT CONSTANTS
appears.
20. Press the Arrow Key to execute. When COMPLETED appears,
open S1-2 to protect the saved settings.
21. Press Key 5 until SYSTEM DISPLAY appears and press the
Arrow Key to select the Operate Mode. This returns the FT-50
to a secured operating mode and completes the calibration
procedure.
22. If you are using this production line to produce more than one
type of container, you must repeat this procedure for each
different container type (i.e. containers with different
diameters, or line speeds).
Industrial Dynamics Company, Ltd. Document 40255-0600