Establishing the Target Head Height
1. Press Key 1 to begin the Learn Target Head Height procedure.
2. Precisely center a 100 percent reject container in the inspection
tunnel and then press any key to continue.
3. Lower the gamma beam into the liquid by slowly turning the
Head Height Adjustment Crank.
4. Continue lowering the head until RAISE HEAD ONE TICK
displays. Stop lowering the head at this point.
5. Raise the head one tick as indicated by the Head Height Meter
and then press the Arrow Key.
6. Repeat step 5 until @TARGET HEAD HEIGHT appears. You
are now at the Target Head Height. Note the head height
indicated in the Head Height Meter, you must enter the height
in the next step.
7. Press Key 1 and then use the Arrow Keys to enter the Target
Head Height into the system.
8. Press Key 1 to accept the height setting and to proceed to the
Dynamic Reject Point Test.