FT-50 Fill Level Inspector Service Manual
9-4
PLC Series Alarms
To configure your FT-50 for operation with a PLC you must
activate the PLC Series Alarms functions in the FT-50’s software.
To activate the PLC Series Alarms:
1. Open dip Switch S1-5 to enter the System Configuration
2. Press Key 5 until SYSTEM DISPLAY appears and press the
Arrow Key to select CALIBRATE.
3. Press Key 5 until PLC SERIES ALARMS appears and press the
Arrow Key to select PRESENT.
4. Close Dip switch S1-5.
The PLC Series Alarm functions are now activated in the system
software. The following options will now appear in the Function
Menus. Refer to Chapter 5: System Software for information on
these functions.
PLC Pin Assignments
PLC Output Pulse Width
Alarm Series Underfills
Alarm Series Overfills
Alarm Series Missing Lid
Alarm Series Bulged End
Alarm Series Down Can
Alarm Series Dia. Missing Lid
Alarm Series Any Other Inspections.