FT-50 Fill Level Inspector Service Manual
4-26
Configure the External Reject Function
Before you activate the External Reject Function, be sure you have
the signal cable wired correctly, see Wiring the External Reject
Function for proper wiring configuration.
1. If your FT-50 is not in the Calibrate Mode, Press Key 5 until you
reach SYSTEM DISPLAY and then press the Arrow Key to
enter the Calibrate Mode.
2. Open dip switch S1-5.
3. Press Key 5 until EXTERNAL REJECT appears and use the
Arrow Key to select PRESENT.
4. Close dip switch S1-5.
5. Press Key 2 until EXTERNAL REJECT STATUS appears and
then press the Arrow Key to activate the function.
6. If you are finished configuring the FT-50, press Key 5 until you
reach SYSTEM DISPLAY and then press the Arrow Key to
return to the Operate Mode.
Industrial Dynamics Company, Ltd. Document 40255-0600