FT-50 Fill Level Inspector Service Manual
8-26
Note
If intermittent or erratic rejector operation continues, contact
Industrial Dynamics Customer Service for assistance.
Containers are rejected erratically.
Time Delay is incorrectly set. Pass reject test container through the
inspector and observe the time delay
synchronization. The rejector pad's centerline
should contact the containers centerline at
the point of impact. If not, adjust the reject
timing settings in the inspection unit.
Rejector Pad fatigue or failure. Fatigued or worn pads cause the rejected
container's trajectory to change. Replace
Rejector Pad.
Facility air pressure drops or is
erratic.
The facility air supply pressure must be
maintained at 10 psi over the regulator
pressure setting. If facility air supply pressure
drops when rejector operation is erratic,
install an accumulator upstream of the
rejector air controls.
Rejector sticking or binding. Check the rejector piston for smooth
movement. If it is binding, you must replace
the Servo unit. See your Proline Manual for
parts information.
Possible Cause Corrective Action
Industrial Dynamics Company, Ltd. Document 40255-0600