19
DMP30 / DMP40
Technical Data
Grouping measured readings
Block (group) size can be set between 1 and 1000 single readings
Storage of measured readings
Can be disabled
Measurement unit
µm, mm, mils, µinch or customized unit (definable, related to µm), selectable
Limit Monitoring
Can be disabled; limit values can be set
Offset value/ correction value
Can be set, is deducted automatically from the measured reading. Thus, one ob-
tains the thickness of the top coating if for instance the interim coating is known.
Resolution of the displayed values
Low (up to 1 decimal place)
Medium (up to 2 decimal places)
High (up to 3 decimal places)
NAVSEA (1 decimal place for measured readings up to 254 µm (10 mils), inte-
ger for measured readings as from 254 µm (10 mils)
Measurement modes
Single Reading
After each placing of the probe, the measured reading is displayed and stored
automatically in the gauge.
Free Running
After placing the probe on the surface, the continuous display of the measured
readings appears without automatic storage, useful for quick checking of coat-
ing thicknesses over a defined surface area, e.g. in tank construction
Scan
The Scan mode allows a defined acquisition of the measured readings while
the probe is moving over the surface. You can set the number of single read-
ings and the time interval for the measurement acquisition in the gauge. Useful
to determine the coating thickness over surface areas.