40
Settings for measurement DMP30 / DMP40
To deactivate automatic block formation
1. > Batch Modify > OK > Block Size > OK > Generate autom. > OK.
Parameter is deactivated:
2. Exit menu: , automatic block formation is deactivated.
7.5 Activating/deactivating offset function
You can correct the measured reading by a pre-defined offset value. The actual mea-
sured reading less the offset value is displayed.
Application example: coating thickness measurement on a 2-coating system with
known intermediate coating thickness that is entered as the offset value. Only the
thickness of the top coating is then displayed.
Before you start
The batch for which you want to make the changes is open in the gauge.
To activate the offset and set the offset value
1. > Batch Modify > OK > Offset > OK > On > OK > set desired offset value > OK
Parameter is activated:
2. Exit the menu: , the settings are now made.
The icon appears in the Measurement view to indicate the active offset func-
tion.
To deactivate the offset function
1. > Batch Modify > OK > Offset > OK > Off > OK
Parameter is deactivated:
2. Exit the menu: , the offset function is deactivated.
[ ]
[ ]