73
DMP30 / DMP40
Calibration
2. Activate the lock function: Lock Calibration > OK.
Function is activated:
3. Exit the menu: , the locking process has finished.
To access the menu and the calibration routine of the locked calibration, you
now need to enter the lock code.
9.13 Calibration – deactivating lock
The Lock Calibration function blocks access to the menu and to the calibration routine
of the selected calibration. You can deactivate the lock function for the menu and the
calibration routine of the locked calibration only by entering the lock code.
Before you start
The probe is connected to the gauge containing the calibration whose lock you
want to deactivate.
To deactivate the lock function of a calibration, you need to know the lock code.
To deactivate the lock of selected calibration
1. Select the calibration whose lock you want to deactivate: > Calibrations > OK
> select name of calibration > OK.
2. Deactivate the lock:
a Set the lock code: set lock code > OK.
b Deactivate the lock function: Lock Calibration > OK.
Function is deactivated:
The unlocking process is finished, you now have free access to the functions in the
menu of the selected calibration.
Before exiting the calibration, do not forget to reactivate the lock function for
the menu of the selected calibration, 72.
[ ]