74
DMP30 / DMP40
Calibration
9.14 Deleting a calibration
You can delete individual calibrations from the memory of the probe/DMP-F probe
adapter.
Before you start
The probe is connected to the gauge containing the memory from which you
want to delete a calibration.
If the calibration you want to delete is locked, you need to know the lock code.
To delete a calibration
1. Select the calibration you want to delete: > Calibrations > OK > select name
of calibration > OK.
2. Prompted to set the lock code?
Yes: Set the lock code and deactivate the lock function, 73.
Function is deactivated:
No: Continue with step 3.
3. Select the delete function: Delete > OK.
4. Start the delete process: .
5. Exit the menu: , the delete process has finished.
9.15 Deleting all calibrations
You can delete all your calibrations from the memory of the connected probe. This
function deletes all the calibration files and all their data!
DMP-F probe adapter: You can use this function to delete all calibrations from the
DMP-F probe adapter that were created using the connected probe. Useful when an
existing probe is to be replaced by a new one. In this case, the calibrations of the old
(e.g. defective) probe are no longer needed, so they can be deleted and do not unnec-
essarily take up memory space in the DMP-F probe adapter.
ATTENTION
Please note that a measurement without a calibration is not possible!
You must assign a new calibration to each batch already created!
Using this function you also delete the locked calibrations in the probe/DMP-
F probe adapter!
[ ]