DMP30 / DMP40 Settings for measurement
7.7 Save measured readings on/off
You can specify whether the measured readings are automatically saved in the open
batch.
Function activated (default mode)
The measured reading is automatically saved in the open batch with every measured
reading acquisition.
Function deactivated
The measured readings are automatically saved in the open batch only temporarily.
When a different batch is called or the gauge is switched off, the measured readings
are lost.
Before you start
The batch for which you want to make the changes is open in the gauge.
To activate/deactivate the save measured reading function
1. > Batch Modify > OK > Save Readings > OK
Parameter is activated:
Parameter is deactivated:
2. Exit the menu: , the setting has now been made.
The Measurement view shows the icon to indicate that the measured read-
ing save function has been deactivated.
7.8 Air reference value automatic/manual acquisition
The air reference value references the zero point determination for the measure-
ment.
Zero point determination: measurement on the uncoated base material = coat-
ing thickness 0 (normalization, zero)
Air reference: Measurement in air = coating thickness infinite = saturation = air
reference value
[ ]