DMP30 / DMP40 Measure
8.7 Measuring with manual recording of the air reference value
If the minimum lift off distance of 4 x the probe measurement range cannot be main-
tained when measuring in cavities and pipes, you can also record the air reference
value manually at regular intervals.
Before you start
Air value recording is set to Static, 43
Refer to the notes on measurement at the beginning of this chapter
The batch is open, the Measurement view can be seen in the display.
To measure with manual recording of the air reference value
1. Perform your measurements as described in the previous chapters.
We recommend recording the air reference value at regular intervals, e.g.
between each specimen or after every 10th specimen.
2. Record the air reference value: > hold the probe in the air > OK > .
The manual recording of the air reference value is now complete.
3. Repeat steps 1. and 2. for subsequent measurements.