64
DMP30 / DMP40
Calibration
Std 2 (NC/NF): 201 μm (example)
Calibration on non-magnetizable base material (NF), indication of the foil thickness
(nominal value) for foil 2 used for this calibration.
Date / Time:
13.04.2022 10:23 (example): storage time of calibration
9.6 Activating/deactivating semi-automatic calibration routine
The Shortened course function activates a semi-automatic calibration routine for the
selected calibration. If the Shortened course function is activated, the calibration rou-
tine switches automatically to the next calibration step after every 3 measurements.
The next calibration step is signaled only through the display of the required stan-
dard value. It is not possible to correct/change the specified calibration standard
when this function is activated.
Before you start
The probe is connected to the gauge containing the calibration for which you
want to activate/deactivate the semi-automatic calibration routine.
If the calibration for which you want to activate or deactivate this function is
locked, you need to know the lock code.
To activate/deactivate the function for a calibration
1. Select the calibration for which you want to activate/deactivate the function:
> Calibrations > OK > select name of calibration > OK.
2. Prompted to set the lock code?
Yes: Set the lock code and deactivate the lock function, 73
function is deactivated:
No: Continue with step 3.
3. Activate/deactivate the function: Shortened course > OK.
Function is activated:
At the next recalibration, the calibration routine will be performed semi-auto-
matically.
Function is deactivated:
[ ]