65
DMP30 / DMP40
Calibration
4. Did you deactivate the lock function in step 2.?
• Yes: Reactivate the lock function, 72
function is activated:
• No: Continue with step 5.
5. Exit the menu: , the function is now activated/deactivated.
9.7 Calibration - reset
You can reset a calibration and delete all the correction values (coating thickness) of
the selected calibration. The function deletes all parameters measured in the Std
calibration steps. The entered standard nominal values and the measurements of
the Zero calibration step are retained.
Before you start
The probe is connected to the gauge containing the calibration you want to re-
set.
If the calibration for which you want to activate or deactivate this function is
locked, you need to know the lock code.
To reset the selected calibration
1. Select the calibration you want to reset: > Calibrations > OK > select name of
calibration > OK.
2. Prompted to set the lock code?
• Yes: Set the lock code and deactivate the lock function, 73.
Function is activated:
• No: Continue with step 3.
3. Start the function: Reset > OK.
4. Confirm the prompt asking whether you want to proceed with the Reset:
5. Did you deactivate the lock function in step 2.?
• Yes: Reactivate the lock function, 72
function is activated:
• No: Continue with step 6.
6. Exit the menu: , the calibration has now been reset.
[ ]