50
Measure DMP30 / DMP40
Measuring on flat specimens - two-tip probe
The display screens should be considered only as examples, the gauge default modes are required here for
the measured reading acquisition, probe D-F90-2t8 is shown as a typical example of a 2-tip probe
Additional measurement with 90° rotated probe when measuring on specimens with
magnetic preferred direction.
Measuring on cylindrical specimens - two-tip probe
The display screens should be considered only as examples, the gauge default modes are required here for
the measured reading acquisition, probe D-F90-2t8 is shown as a typical example of a 2-tip probe
1. 2. 3.
Position the gauge
Measured reading ac-
quisition
Lift off the probe
≥ 4 x probe
measurement
range
No.
x
Min
n
s
Max
1
119
118
2
1
120
NF/FE
Batch 003
BLOCK
115
120
125
µm
No.
x
Min
n
s
Max
2
122
122
1
---
122
NF/FE
Batch 003
BLOCK
115
122
125
µm
Coated specimen
No.
x
Min
n
s
Max
2
122
122
1
---
122
NF/FE
Batch 003
BLOCK
115
122
125
µm
Coated specimen
90°
4. 6.
Position the
probe
Acquire the
measured
reading
Rotate probe
through 90°
7.
Lift off probe
5.
No.
x
Min
n
s
Max
2
122
122
1
---
122
NF/FE
Batch 003
BLOCK
115
122
125
µm
No.
x
Min
n
s
Max
2
122.5
122
2
0.25
123
NF/FE
Batch 003
BLOCK
115
123
125
µm
No.
x
Min
n
s
Max
2
122.5
122
2
0.25
123
NF/FE
Batch 003
BLOCK
115
123
125
µm
1. 2. 3.
Position the gauge
Measured reading ac-
quisition
Lift off the probe
x ≥ 4 x probe mea-
surement range
Coated
specimen
x
No.
x
Min
n
s
Max
1
121
118
4
1.2
123
NF/FE
Batch 001
BLOCK
115
120
125
µm
No.
x
Min
n
s
Max
1
121.5
118
5
1.3
123
NF/FE
Batch 001
BLOCK
115
123
125
µm
No.
x
Min
n
s
Max
1
121.5
118
5
1.3
123
NF/FE
Batch 001
BLOCK
115
123
125
µm