58
Calibration DMP30 / DMP40
Before you start
Refer to the notes on calibration, 56
The probe in which the calibration to be recalibrated is stored is connected to
the gauge.
Reference piece and/or calibration foils are ready. You can find the parts you
need on the calibration info page: > Calibrations > OK > select name of cali-
bration > OK > Info.
If the calibration you want to repeat is locked, you need to know the lock code.
To repeat an existing calibration
1. Open the calibration to be repeated:
Open a calibration from the list: > Calibrations > OK > select name of cali-
bration > OK > CALIBRATE.
Open the calibration used in the open batch: (in the Measurement view).
2. Prompted to set the lock code?
Yes: Set the lock code and deactivate the lock function, 73
function is deactivated:
No: Continue with step 3.
3. Follow the routine and perform the calibration steps (Zero, Std. 1/2).
a "Calibration step Zero Procedure", 59
b "Calibration step Std – procedure", 61
Notes
If the Shortened course function is activated, the calibration routine switches
automatically to the next calibration step after every 3 measurements. It is
not possible to correct/change the specified calibration foil when this func-
tion is activated.
If you have deactivated the lock function for recalibration, do not forget to
reactivate the lock function before exiting the calibration menu, 72.
[ ]
[ ]