ActionStep
Place the tube rack (9) over the central spindle (11) and pull the spring loaded
drive sleeve (12) out so the tube rack comes to rest.
6
Insert collection tubes
Insert a sufficient number of collection tubes in to the tube rack, starting at position 1,
pushing each one down as far as they will go. All the tubes must be of the same length
and diameter and there should be no spaces in the sequence.
Adjust the delivery arm
Follow the instructions to adjust the height of the delivery arm.
NOTICE
Never lift Fraction collector F9-R by the delivery arm. This may
damage the fraction collector.
214 ÄKTA avant User Manual 29035184 AE
5 Operation
5.8 Fractionation
5.8.2 Prepare Fraction collector F9-R