Possible cause and actionProblem
The air intake on the rear or on the right side of the
instrument is covered
Make sure that none of the air intakes on the instru-
ment are covered.
Hot surroundings
Decrease the room temperature. Maximum operating
temperature is 35°C.
Hardware error
Switch off the instrument and wait until the tempera-
ture has decreased. Restart the instrument. If this error
is recurrent, generate a System error report and con-
tact Service.
The internal tempera-
ture of the UV monitor
is too high
Wrong UV flow cell for current buffer
Change to a shorter UV flow cell or change buffer.
The UV flow cell is not correctly installed
Check that the UV flow cell is fitted correctly, see Sec-
tion 7.8.5 Replace UV monitor U9-L flow cell, on
page 345.
Dirt in the UV flow cell
Clean the UV flow cell, see Section 7.5.1 Clean the UV
flow cell, on page 283.
Broken UV flow cell
Replace the cell, see Section 7.8.5 Replace UV monitor
U9-L flow cell, on page 345.
Maximum absorbance
that can be measured
by the detector is
reached
Check that the UV flow cell is fitted correctly, see Sec-
tion 7.8.5 Replace UV monitor U9-L flow cell, on
page 345.
The UV cell is not cor-
rectly installed
ÄKTA avant User Manual 29035184 AE 369
8 Troubleshooting
8.3 Troubleshooting: Monitors