Inlet valves
Possible cause and actionProblem
Hardware error
Restart the instrument with the power switch. If this
error is recurrent, generate a System error report and
contact Service.
Faulty air sensor in the
valve
pH valve
Possible cause and actionProblem
The dummy electrode was dry when it was installed
in the valve.
1
Remove the dummy electrode.
2
Wet the dummy electrode properly with distilled
water.
3
Insert the dummy electrode into the pH valve.
4
Rotate the dummy electrode before securing it
with the nut.
Leaking pH valve
Module Panel
Possible cause and actionProblem
A Module Panel is missing and the position is left
empty
Install the missing Module Panel.
The cable between the Module Panel and the ICU is
not connected
Remove the Module Panel and make sure that the
cable is connected.
The instrument is un-
able to find some of the
modules
382 ÄKTA avant User Manual 29035184 AE
8 Troubleshooting
8.4 Troubleshooting: Valves