Quaternary valve
Possible cause and actionProblem
The sum of the flow from inlet Q1-Q4 is not equal
to 100% of the total flow
Make sure that the sum of the flow from inlet Q1-Q4
is equal to 100% of the total flow.
Wrong mixing ratio for
Quaternary valve
The switch valve is not opened and closed correctly
Perform Performance test according to ÄKTA avant
Unpacking Instructions. Check result and, if necessary,
generate a System error report and contact Service.
Unexpected values for
e.g., pH curve or con-
ductivity curve using a
quaternary flow
ÄKTA avant User Manual 29035184 AE 383
8 Troubleshooting
8.4 Troubleshooting: Valves