Possible cause and actionProblem
Quick scan has not detected the correct Cassette
Clean the Cassette type code of the Cassette. See
Section 7.6.4 Clean built-in fraction collector, on
page 301.
Unsupported deep well plate is used.
Make sure that approved deep well plates are used.
See Section 3.10.3 Cassettes, cassette tray and racks,
on page 89.
Wells in deep well plates are prefilled to a volume
above 25% of the total well volume
Full scan will not work with prefilled wells during these
conditions.
The deep well plate is not correctly positioned in
the Cassette
See Section 5.8.1 Prepare the built-in fraction collector,
on page 203 for information of how to place the deep
well plate
Dirt on the nozzle or DropSync sensor diode win-
dows. Dirt may effect where the drops fall.
Clean nozzle and DropSync sensor diode windows.
See Section 7.6.4 Clean built-in fraction collector, on
page 301.
The Fraction collector
fractionates in the
wrong well or tube
Some parts in a method require that the Frac
drawer is closed.
Do not open the Frac drawer when the symbol indicat-
ing ongoing fractionation is displayed on the Instru-
ment display (see Buttons and indicators, on page41).
Error message when
the Frac drawer is
opened during a run
The Detector - Outlet valve delay volume has been
set to zero.
1
In System Control, select System:Settings:Frac-
tion collections:Delay volumes.
2
Correct the Detector - Outlet valve delay volume.
The instrument does
not fractionate via the
Outlet valve ports
392 ÄKTA avant User Manual 29035184 AE
8 Troubleshooting
8.5 Troubleshooting: Fraction collector