Volume (ml)Length
(mm)
TubingDescriptionLabel
ÄKTA avant
150
ÄKTA avant
25
ÄKTA avant
150
ÄKTA avant
25
6.62.01000FEP, o.d.
3/16", i.d. 2.9
mm
FEP, o.d.
1/8", i.d. 1.6
mm
Inlets to Sample
inlet valve
S1 to
S7
1
9.93.01500FEP, o.d.
1/8", i.d. 1.6
mm
FEP, o.d.
1/8", i.d. 1.6
mm
Inlets to Quate-
nary valve
Q1 to
Q4
9.93.01500FEP, o.d.
1/8", i.d. 1.6
mm
FEP, o.d.
1/8", i.d. 1.6
mm
Quatenary valve
to Inlet valve A
InQA
9.93.01500FEP, o.d.
1/8", i.d. 1.6
mm
FEP, o.d.
1/8", i.d. 1.6
mm
Quatenary valve
to Inlet valve B
InQB
6.62.01000FEP, o.d.
3/16", i.d. 2.9
mm
FEP, o.d.
1/8", i.d. 1.6
mm
Inlet to Sample in-
let valve
Buff
3.81.2580FEP, o.d.
3/16", i.d. 2.9
mm
FEP, o.d.
1/8", i.d. 1.6
mm
Sample inlet valve
to Sample pump
InS
1
Narrow inlet tubing, ETFE, o.d. 1/16", i.d. 0.75 mm (28-9572-17), is available for S1-S7.
High pressure tubing
The tables below shows the labels, standard diameters, and standard lengths of the
standard high pressure tubing and the optional high pressure tubing.
Standard high pressure tubing
Volume (μl)Length
(mm)
TubingDescriptionLabel
ÄKTA avant
150
ÄKTA avant
25
ÄKTA avant
150
ÄKTA avant
25
270150340PEEK, o.d.
1/16", i.d. 1.0
mm
PEEK, o.d.
1/16", i.d.
0.75 mm
System pump A
left to Restrictor A
1A1
ÄKTA avant User Manual 29035184 AE 435
9 Reference information
9.3 Tubing and connectors