Use with tubing...Description
ETFE, o.d. 1/16", i.d. 0.75 mmTubing connector 5/16" +
Ferrule (blue), 1/16"
FEP o.d. 3/16" i.d. 2.9 mmTubing connector 5/16" + ferrule
(blue) 3/16"
FEP, o.d. 1/8", i.d. 1.6 mmTubing connector 5/16" +
Ferrule (yellow), 1/8"
Other connectors
The table below shows other connectors used in ÄKTA avant.
Scope of useDescription
Stop plug for valve portsStop plug 1/16" male
Syringe connector for pH valve and Injection valveLuer female to 1/16" male
Inlet tubing
The table below shows the labels, standard diameters, and standard lengths of the inlet
tubing.
Volume (ml)Length
(mm)
TubingDescriptionLabel
ÄKTA avant
150
ÄKTA avant
25
ÄKTA avant
150
ÄKTA avant
25
9.93.01500FEP, o.d.
3/16", i.d. 2.9
mm
FEP, o.d.
1/8", i.d. 1.6
mm
Inlets to Inlet
valve A
A1 to
A7
9.93.01500FEP, o.d.
3/16", i.d. 2.9
mm
FEP, o.d.
1/8", i.d. 1.6
mm
Inlets to Inlet
valve B
B1 to
B7
2.00.6300FEP, o.d.
3/16", i.d. 2.9
mm
FEP, o.d.
1/8", i.d. 1.6
mm
Inlet valve A to
System pump A
InA
2.00.6300FEP, o.d.
3/16", i.d. 2.9
mm
FEP, o.d.
1/8", i.d. 1.6
mm
Inlet valve B to
System pump B
InB
434 ÄKTA avant User Manual 29035184 AE